Overview of a Verigene® Test Cartridge

Non flash


trans

General description:

  • Single use, self-contained, test unit comprised of the following:
    • Reagent pack — single use plastic carrier that creates a hybridization chamber on the surface of the array, houses all of the reagents needed for a single Test, and captures all of the waste materials generated during Test processing.
    • Slide (an array) — glass slide, solid support for the capture oligonucleotides required by the Test.

Capacity:

  • Each Test Cartridge is designed for single-use analysis of one sample.
  • The typical sample loading volume is 25 µL.

Functionality:

  • All of the wet chemistry required to run a test occur inside a Test Cartridge, including:
    • Fragmentation and denaturation of genomic DNA or RNA via on-board sonication.
    • Primary hybridization of genomic DNA or RNA to capture probe oligonucleotides and mediator oligonucleotides under conditions of sufficient stringency. 
    • Washing away of unbound sample nucleic acids and mediator oligonucleotides.
    • Secondary hybridization of gold nanoparticle probes to captured target nucleic acid / mediator oligonucleotide complexes.
    • Washing away of unbound gold nanoparticle probes with oligonucleotides.
    • Signal amplification of gold nanoparticles oligonucleotide probes.
    • Washing away of unreacted signal amplification reagents.
FDA cleared for in vitro diagnostic (IVD) with specific Verigene® Tests.  Please refer to the Verigene® Nucleic Acid Tests section.